Helena Hemostasis Coagulation
Helena Hemostasis - Coagulation

Manual Coagulation Analyzers

Cascade M and Cascade M-4 are manual hemostasis analyzers that are ideal for all routine clotting assays - PTs, aPTTs, thrombins, fibrinogens and factor assays. Designed for easy, inexpensive, low-volume testing, these analyzers are small in size but big on features, offering you many time and labor saving benefits. Operation is simple and economical.

 


Cascade M and Cascade M-4

Cat. No. Item
1710 Cascade M, 2-channel analyzer, 4-line screen, 110V / 220V
1711 Cascade M-4, 4-channel analyzer, 6-line screen, 110V / 220V
1194 Comprehensive Replacement Agreement for Cascade M
1195 Comprehensive Replacement Agreement for Cascade M-4
9164 Service Manual for Cascade M
9167 Service Manual for Cascade M-4


PT, aPTT, Fibrinogen, Thrombin Clotting Time, Factor AssaysEasy-to-Use Keypad, On-Screen Prompts - All tests are clearly enumerated on the keypad for fast, easy test selection - there's no guess work caused by "hidden" function keys. Computer prompts appear on the screen to conveniently step the user through testing. Default settings are preprogrammed but can easily be changed, if desired. (Instructions for changing parameters are provided).

State-of-the-Art Photo Optics - Independent photo-optic detection sources (two for Cascade M, four for Cascade M-4) read all samples, regardless of turbidity. Samples can be read in duplicate or in single.

PT, aPTT, Fibrinogen, Thrombin Clotting Time, Factor AssaysComputerized Features, On-board Quality Control - Cascade M and M-4 store standard curves, calculate patient results, INRs and aPTT ratios. Quality control tasks are easy because Cascade M and M-4 can calculate mean, SD and CV, plus plot and print Levy-Jennings QC charts.

Compact, Lightweight - With benchtop footprints smaller than 15" wide by 15" deep and weights of approximately 20 pounds, Cascade M and M-4 are sized to fit any lab. Compact size, dependability and versatility make Cascade M and M-4 excellent choices as primary analyzers for smaller laboratories or as backup units for larger coagulation analyzers.


Cascade M & Cascade M-4 Specifications

Tests: PT, aPTT, Fibrinogen, Thrombin Clotting Time, Factor Assays.
Sample Size: 50 μL, 100 μL, 200 μL.
Optics System: Constant current LED photo optic detection.
Channels: 2 for Cascade M; 4 for Cascade M-4.
Temperature Control: 37°C ± 0.5°C.
Special Features: Stores standard curves; calculates final results for fibrinogens and factor assays. Calculates INRs and aPTT ratios. On-screen prompts for test instructions. LIS interfacing (Cascade M-4).
Quality Control Features: Calculates mean, SD, and CV. Plots Levy-Jennings charts.
Display: Fluorescent LCD. Cascade M: 4 lines x 40 characters. Cascade M-4: 6 lines x 40 characters.
Printer: Accommodates 4 1/2" thermal paper. Reports include date, time, INR, average, technician and patient ID. Prints QC and standard curves.
Power Requirements: 110/220 Vac, 50/60 Hz, 100 Watts max.
Operating Environment: Room temperature.
Dimensions: Cascade M: 8" H x 15" W x 14" D (20 cm H x 38 cm W x 35 cm D).
Cascade M-4: 8 1/2" H x 14" W x 13" D (22 cm H x 35 cm W x 33 cm D).
Weight: Cascade M: 20 lbs. (9 kg). Cascade M-4: 21 lbs. (9.5 kg).

Cascade M-4 accessoriesCascade M and Cascade M-4 Accessories

Cat. No. Item
1718 Cascade M / M-4 Cuvettes, 5 x 100/pkg
1720 Cascade M / M-4 Printer Paper, 5 rolls
1475 Disposable Pipette Tips (50-200 µL), 1000 pkg

 

 

THOR Accessories, Replacement Parts, Service Agreements

Cat. No. Item
1465 Toner Cartridge for Brother HL-630 Laser Printer, 1 pkg
9159 Service Manual for THOR

 

CSA-1200 Accessories

Cat. No. Item
1442 Semi-Micro Cuvettes for CSA-1200, 500 pkg